How Simoa counts single protein molecules
A sandwich immunoassay, like ELISA, read in femtoliter microwells. At low occupancy, most beads carry zero or one labeled complex. Confining the enzymatic product makes it possible to count positive wells and detect very low concentrations. The limit of detection depends on the analyte, background and assay validation.
Same enzyme, a billion times less volume
From plasma to a fluorescence image
Step through the assay. The first four steps happen in suspension on paramagnetic beads (about 2.7 µm, roughly 500,000 per test); the last two happen on the array disc.
Counting wells, and why Poisson matters
Because there are far more beads than target molecules at low concentration, each bead carries either zero or one enzyme most of the time. The number of enzymes per bead follows a Poisson distribution with mean λ. The instrument measures the fraction of bead-containing wells that are fluorescent, fon, and converts it to average enzymes per bead (AEB).
analog: AEB = fon × Īon / Īsingle (used at higher load, where counting saturates)
Move the concentration slider and watch the array, the occupancy histogram, and where you sit on the calibration curve. A small nonspecific background is included, which sets the floor that defines the limit of detection.
Enzymes per bead (Poisson, share of beads)
Model assumptions, chosen to be in a realistic range: 100 µL effective sample, 500,000 beads, 50% capture efficiency, about 60% of wells receiving a bead. Real assays differ by analyte, antibody pair and dilution, so read the axis as illustrative, not as a specification.
Shrink the volume, keep the enzyme
One streptavidin β-galactosidase molecule turning over substrate for 30 seconds makes on the order of 104 resorufin molecules (assuming ~300 turnovers per second). What that means as a concentration depends only on the volume it is trapped in.
What to keep in mind when reading Simoa data
Calibration is per analyte
AEB is converted to concentration with a calibrator curve (typically a 4PL fit). Lot changes in beads, detector antibody or calibrator can shift absolute values, so bridging and internal controls matter in longitudinal cohorts.
Matrix still matters
Single-molecule counting removes the volume problem, not the biology. Heterophilic antibodies, hemolysis, freeze–thaw history and sample dilution all act before the bead reaches the well.
Two-step or three-step
Some kits add capture and detector antibodies together with sample; others separate the steps. Format affects hook effect risk, incubation time and sensitivity, and is defined in each kit's protocol.